Getting Started
R&S
®
FSW
78User Manual 1173.9411.02 ─ 19
<NOR | APR>
Ext. Gen
An external generator is being controlled by the R&S FSW (requires option R&S FSW-
B10).
NOR: the measurements are normalized with the results of the external generator cali-
bration
APR (approximation): the measurements are normalized with the results of the external
generator calibration; however, the measurement settings have been changed since cali-
bration
If neither label is displayed, no calibration has been performed yet or normalization is not
active.
For details see chapter 6.2.4, "External Generator Control (Option R&S FSW-B10)",
on page 331.
LVL A level offset is applied to the external generator signal (only if external generator control
is active).
3.4.1.2 Window Title Bar
Each channel in the R&S FSW display may contain several windows. Each window
can display either a graph or a table as a result of the channel measurement. Which
type of result evaluation is displayed in which window is defined in the display configu-
ration (see chapter 3.4.5, "Displaying Results", on page 85). The window's title bar
indicates which type of evaluation is displayed.
Trace Information in Window Title Bar
Information on the displayed traces is indicated in the window title bar.
Trace color
Color of trace display in diagram
Trace no.
Trace number (1 to 6)
Detector
Selected detector:
AP AUTOPEAK detector
Pk MAX PEAK detector
Mi MIN PEAK detector
Sa SAMPLE detector
Av AVERAGE detector
Rm RMS detector
QP QUASIPEAK detector
Trace Mode
Sweep mode:
Clrw CLEAR/WRITE
Operating the Instrument