Remote Commands
R&S
®
FSW
784User Manual 1173.9411.02 ─ 19
CALCulate<n>:STATistics:SCALe:Y:LOWer <Magnitude>
This command defines the lower vertical limit of the diagram.
Parameters:
<Magnitude> The number is a statistical value and therefore dimensionless.
Range: 1E-9 to 0.1
*RST: 1E-6
Example:
CALC:STAT:SCAL:Y:LOW 0.001
Manual operation: See "Y-Max / Y-Min" on page 260
CALCulate<n>:STATistics:SCALe:Y:UNIT <Unit>
This command selects the unit of the y-axis.
Parameters:
<Unit> PCT | ABS
*RST: ABS
Example:
CALC:STAT:SCAL:Y:UNIT PCT
Sets the percentage scale.
Manual operation: See "Y-Unit" on page 260
CALCulate<n>:STATistics:SCALe:Y:UPPer <Magnitude>
This command defines the upper vertical limit of the diagram.
Parameters:
<Magnitude> The number is a statistical value and therefore dimensionless.
Range: 1E-5 to 1.0
*RST: 1.0
Example:
CALC:STAT:SCAL:Y:UPP 0.01
Manual operation: See "Y-Max / Y-Min" on page 260
11.5.8.5 Performing a Statistical Measurement
The following commands are required to perform a statistical measurement:
INITiate[:IMMediate] on page 694, see chapter 11.5.1, "Performing Measure-
ments", on page 692
11.5.8.6 Retrieving Results
The following commands are required to retrieve the measurement results.
Useful commands for retrieving results described elsewhere:
CALCulate<n>:MARKer<m>:X on page 932
Configuring and Performing Measurements