Remote Commands
R&S
®
FSW
815User Manual 1173.9411.02 ─ 19
<TriggerSource> EXTernal | EXT2 | EXT3 | IMMediate | IFPower | RFPower |
VIDeo
Selects a trigger source.
For more information see Configuring Triggered and Gated
Measurements.
<TriggerSlope> NEGative | POSitive
Selects the trigger slop.
<TriggerOffset> Defines the trigger delay.
Range: negative measurement time to 30
*RST: 0
Default unit: s
<GateLength> Defines the gate length for gated measurements.
Setting 0 seconds turns gated measurements off.
To perform gated measurements, the trigger source must be dif-
ferent from IMMediate.
Range: 31.25 ns to 30 s
*RST: 0 s
[SENSe:]LIST:POWer:STATe <State>
This command turns the List Evaluation off.
Parameters:
<State> OFF
*RST: OFF
11.5.14.2 Example: Performing List Evaluation
The following example shows a list evaluation with the following configuration.
No Freq
[MHz]
Ref
Level
[dBm]
RF
Attenu-
ation
[dB]
El
Attenu-
ation
[dB]
Filter RBW VBW Meas
Time
Trigger
Level
1 935.2 0 10 --- Normal 1 MHz 3 MHz 440 µs 0
2 935.4 0 10 10 Channel 30 kHz 100 kHz 440 µs 0
3 935.6 0 10 20 Channel 30 kHz 100 kHz 440 µs 0
-----Measurement with synchronization via service request-----
*ESE 1
*SRE 32
// Configures the status reporting system to produce a service request.
LIST:POW:SET ON,ON,OFF,EXT,POS,10us,434us
//Turns on the list evaluation, configures the global list evaluation settings and
//evaluates the peak and RMS power.
Configuring and Performing Measurements