Remote Commands
R&S
®
FSW
833User Manual 1173.9411.02 ─ 19
[SENSe:]FREQuency:CENTer:STEP:LINK:FACTor <Factor>
This command defines a step size factor if the center frequency step size is coupled to
the span or the resolution bandwidth.
Parameters:
<Factor> 1 to 100 PCT
*RST: 10
Example:
FREQ:CENT:STEP:LINK:FACT 20PCT
Manual operation: See "Center Frequency Stepsize" on page 390
[SENSe:]FREQuency:OFFSet <Offset>
This command defines a frequency offset.
If this value is not 0 Hz, the application assumes that the input signal was frequency
shifted outside the application. All results of type "frequency" will be corrected for this
shift numerically by the application.
See also "Frequency Offset" on page 390.
Parameters:
<Offset> Range: -100 GHz to 100 GHz
*RST: 0 Hz
Example:
FREQ:OFFS 1GHZ
Usage: SCPI confirmed
Manual operation: See "Frequency Offset" on page 390
[SENSe:]FREQuency:SPAN <Span>
This command defines the frequency span.
If you set a span of 0 Hz in the Spectrum application, the R&S FSW starts a measure-
ment in the time domain.
Parameters:
<Span> The minimum span for measurements in the frequency domain
is 10 Hz. For SEM and spurious emission measurements, the
minimum span is 20 Hz.
Range: 0 Hz to fmax
*RST: Full span
Usage: SCPI confirmed
Manual operation: See "Zero Span" on page 112
See "Span" on page 388
See "Last Span" on page 390