Remote Commands
R&S
®
FSW
839User Manual 1173.9411.02 ─ 19
Manual operation: See "VBW" on page 410
See "RBW/VBW " on page 411
See "Default Coupling" on page 412
[SENSe:]BANDwidth|BWIDth:VIDeo:RATio <Ratio>
This command defines the coupling ratio of the video bandwidth to the resolution band-
width (RBW/VBW).
Parameters:
<Ratio> Range: 0,001 to 1000
*RST: 1
Example:
BAND:VID:RAT 3
Sets the coupling of video bandwidth to video bandwidth =
3*resolution bandwidth
Usage: SCPI confirmed
Manual operation: See "RBW/VBW " on page 411
[SENSe:]BANDwidth|BWIDth:VIDeo:TYPE <Mode>
This command selects the position of the video filter in the signal path.
Changing the video filter position is possible only if the resolution bandwidth is
100 kHz.
Parameters:
<Mode> LINear
The video filter is applied in front of the logarithmic amplifier.
In linear mode, measurements with a logarithmic level scale
result in flatter falling edges compared to logarithmic mode. The
reason is the conversion of linear power values into logarithmic
level values: if you halve the linear power, the logarithmic level
decreases by 3 dB.
LOGarithmic
The video filter is applied after the logarithmic amplifier
*RST: LINear
Example:
BAND:VID:TYPE LIN
Video filter ahead of the logarithmic amplifier
11.7.2.2 Configuring the Sweep
Useful commands for configuring sweeps described elsewhere:
[SENSe:]AVERage:COUNt on page 911
[SENSe:]AVERage<n>[:STATe<t>] on page 912
[SENSe:]AVERage<n>:TYPE on page 912
Setting Basic Measurement Parameters