Remote Commands
R&S
®
FSW
886User Manual 1173.9411.02 ─ 19
Manual operation: See "(Automatic) Source Frequency (Numerator/Denominator/
Offset)" on page 346
SOURce:EXTernal:POWer[:LEVel] <Level>
This command sets the output power of the selected generator.
Parameters:
<Level> <numeric value>
*RST: -20 dBm
Example:
SOUR:EXT:POW -30dBm
Sets the generator level to -30 dBm
Manual operation: See "Source Power" on page 345
SOURce:EXTernal[:STATe] <State>
This command activates or deactivates the connected external generator.
Parameters:
<State> ON | OFF
*RST: OFF
Manual operation: See "Source State" on page 345
SOURce:POWer[:LEVel][:IMMediate]:OFFSet <Offset>
This command defines a level offset for the external generator level. Thus, for exam-
ple, attenuators or amplifiers at the output of the external generator can be taken into
account for the setting.
Parameters:
<Offset> Range: -200 dB to +200 dB
*RST: 0dB
Example:
SOUR:POW:OFFS -10dB
Sets the level offset of the external generator to – 20 dBm.
Usage: SCPI confirmed
Manual operation: See "Source Offset" on page 345
Interface Configuration
The following commands are required to configure the interface for the connection to
the external generator.
SOURce:EXTernal:ROSCillator[:SOURce]........................................................................887
SYSTem:COMMunicate:GPIB:RDEVice:GENerator:ADDRess............................................887
SYSTem:COMMunicate:RDEVice:GENerator:INTerface.................................................... 887
Setting Basic Measurement Parameters