Remote Commands
R&S
®
FSW
896User Manual 1173.9411.02 ─ 19
Example:
CAL:PMET2:ZERO:AUTO ONCE;*WAI
Starts zeroing the power sensor 2 and delays the execution of
further commands until zeroing is concluded.
Usage: Event
Manual operation: See "Zeroing Power Sensor" on page 325
CALCulate<n>:PMETer<p>:RELative[:MAGNitude] <RefValue>
This command defines the reference value for relative measurements.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<RefValue> Range: -200 dBm to 200 dBm
*RST: 0
Example:
CALC:PMET2:REL -30
Sets the reference value for relative measurements to -30 dBm
for power sensor 2.
Manual operation: See "Reference Value" on page 327
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]:AUTO ONCE
This command sets the current measurement result as the reference level for relative
measurements.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
ONCE
Example:
CALC:PMET2:REL:AUTO ONCE
Takes the current measurement value as reference value for rel-
ative measurements for power sensor 2.
Usage: Event
Manual operation: See "Setting the Reference Level from the Measurement (Meas-
>Ref)" on page 327
CALCulate<n>:PMETer<p>:RELative:STATe <State>
This command turns relative power sensor measurements on and off.
Suffix:
<p>
.
1...4
Power sensor index