Remote Commands
R&S
®
FSW
915User Manual 1173.9411.02 ─ 19
Parameters:
<State> ON | OFF
*RST: OFF
Example:
INIT:CONT OFF
Selects single sweep mode.
INIT;*WAI
Starts the sweep and waits for the end of the sweep.
CALC:SGR:CONT ON
Repeats the single sweep measurement without deleting the
results of the last measurement.
Manual operation: See "Single Sweep/ RUN SINGLE" on page 414
See "Continue Frame" on page 416
CALCulate:
SGRam|SPECtrogram:FRAMe:COUNt <Frames>
This command defines the number of frames to be recorded in a single sweep.
Parameters:
<Frames> The maximum number of frames depends on the history depth.
Range: 1 to history depth
Increment: 1
*RST: 1
Example:
INIT:CONT OFF
Selects single sweep mode.
CALC:SGR:FRAM:COUN 200
Sets the number of frames to 200.
Manual operation: See "Frame Count" on page 416
CALCulate:SGRam|SPECtrogram:FRAMe:SELect <Frame> | <Time>
This command selects a specific frame for further analysis.
The command is available if no measurement is running or after a single sweep has
ended.
Parameters:
<Frame> Selects a frame directly by the frame number. Valid if the time
stamp is off.
The range depends on the history depth.
<Time> Selects a frame via its time stamp. Valid if the time stamp is on.
The number is the distance to frame 0 in seconds. The range
depends on the history depth.
Example:
INIT:CONT OFF
Stop the continuous sweep.
CALC:SGR:FRAM:SEL -25
Selects frame number -25.
Analyzing Measurements (Basics)