Remote Commands
R&S
®
FSW
930User Manual 1173.9411.02 ─ 19
Parameters:
<Mode> ABSolute
Delta marker position in absolute terms.
RELative
Delta marker position in relation to a reference marker.
*RST: RELative
Example:
CALC:DELT:MODE ABS
Absolute delta marker position.
CALCulate<n>:DELTamarker<m>:MREF <Reference>
This command selects a reference marker for a delta marker other than marker 1.
The reference may be another marker or the fixed reference.
Parameters:
<Reference> 1 to 16
Selects markers 1 to 16 as the reference.
FIXed
Selects the fixed reference as the reference.
Example:
CALC:DELT3:MREF 2
Specifies that the values of delta marker 3 are relative to marker
2.
Manual operation: See "Reference Marker" on page 300
CALCulate<n>:DELTamarker<m>[:STATe] <State>
This command turns delta markers on and off.
If necessary, the command activates the delta marker first.
No suffix at DELTamarker turns on delta marker 1.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:DELT2 ON
Turns on delta marker 2.
Manual operation: See "Marker State" on page 299
See "Marker Type" on page 300
CALCulate<n>:DELTamarker<m>:TRACe <Trace>
This command selects the trace a delta marker is positioned on.
Note that the corresponding trace must have a trace mode other than "Blank".
If necessary, the command activates the marker first.