Remote Commands
R&S
®
FSW
964User Manual 1173.9411.02 ─ 19
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK ON
Switches on marker 1.
CALC:MARK:COUN ON
Switches on the frequency counter for marker 1.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:COUN:FREQ?
Outputs the measured value.
Manual operation: See "Signal Count Marker State" on page 508
CALCulate<n>:MARKer<m>:COUNt:FREQuency?
This command queries the frequency at the marker position.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps. See also INITiate:CONTinuous on page 694.
Before you can use the command, you have to turn on the frequency counter.
Return values:
<Frequency> Frequency at the marker position.
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK ON
Switches on marker 2.
CALC:MARK:COUN ON
Activates the frequency counter for marker 1.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:COUN:FREQ?
Outputs the measured value of marker 1.
Usage: Query only
Manual operation: See "Signal Count Marker State" on page 508
CALCulate<n>:MARKer<m>:COUNt:RESolution <Resolution>
This command defines the resolution of the frequency counter.
Parameters:
<Resolution> 0.001 | 0.01 | 0.1 | 1 | 10 | 100 | 1000 | 10000 Hz
*RST: 0.1 Hz
Example:
CALC:MARK:COUN:RES 1kHz
Sets the resolution of the frequency counter to 1 kHz.
Manual operation: See "Resolution" on page 508