6 Maintenance
Attention!
Work on the Power Supply NGMO must be performed in strict
compliance with the ESD-regulations (electrostatic discharge).
6.1 General
In order to maintain the function and the durability of the Power Supply
NGMO, care, maintenance and inspection tasks must be performed at
specified intervals.
6.2 Maintenance Intervals
The following table shows the different activities concerning care,
maintenance and inspection to be performed on the Power Supply
NGMO at the specified intervals.
Maintenance interval
Maintenance tasks
Service point
depending on degree of
soiling
• clean
Unit
Visual inspection before
start-up
• Check connecting cables incl. connections for
damage and tight fit.
Cables and
connections
2 years
• Check the function of the fan
Fan
6.3 Calibration Intervals
Calibration interval
Calibration
Calibration
point
12 months
• NGMO (see chapter 7)
See page 2-1
3.00 / 05-2014 6-1