8.1.3 Setting the NGMO Device Name (from firmware ver-
sion 1.10)
Function test The device name is stored in the EEPROM of the Power Supply
NGMO.
Read out the serial number with the following remote command:
*IDN?
Fault rectification In case of an occurring fault:
Reprogram the device name.
Setting, input After replacing the processor board the device name must be repro-
grammed, if the name displayed when switching on does not match
the designation on the front panel.
Enter the device name using the following remote command:
rus:service:set:ngmo:name “NGMO-NAME"
Instead of “NGMO-NAME“, enter the device name (e.g. “NGMO2“),
as printed on the front panel of the unit.
Quotation marks (") must also be entered.