Repair
P
o
w
S
upp
l
y
NG
M
O
Function test Set any output voltages on channels A and B. Switch both outputs to
ON (see section 5.1).
Switch on test 5 with the following remote command:
Rus:service:select:function:test 5
The display shows the following message:
Inhibit ON test
all outputs OFF
Interrogate test 5:
rus:service:select:function:test?
The return value of this function is always 0.
On both outputs A and B 0 V must be applied to Force +/Force -
(see Fig. 8-2).
Fault rectification
In case of an occurring fault:
Replace the analog board as described in chapter 9 or
replace the processor board as described in chapter 9.
Check the calibration adapter.
3.00 / 05-2014 8-13