Getting started
R&S
®
TSME6
22User Manual 4900.0033.02 ─ 08
Dedicated LAN adapter and IP address for host PC
It is important for the host PC to have its own dedicated LAN adapter for the connec-
tion to one or more R&S TSME6s (or a switch), rather than being integrated in a regu-
lar office network.
If multiple R&S TSME6s are connected to one host PC, the following rules are valid.
Using "static IP" or "partial auto-IP" mode, it is important to define unique IP
addresses for each instrument using the R&S TSME Device Manager (see
R&S TSME6 user manual, chapter Configuring the R&S TSME6).
Using "full auto-IP" mode, it is recommended to define unique IP addresses for
each instrument, but depending on the software run, it may not be mandatory.
Connect the supplied LAN cable to the LAN connector on the rear panel of the
R&S TSME6, and to the host PC.
Windows 10 automatically detects the network connection and all devices in the
same subnet when the R&S TSME6 is switched on.
3.1.5.2 Connecting external devices
The SMA connector is sensitive to mechanical stress. Use the following handling pre-
cautions.
Always use a torque wrench and mount the cable end with 60 Ncm to 90 Ncm.
Do not stack adapters directly at the SMA connector. If you need to use adapters
(e.g: SMA to N), then always use a specific adapter cable (order no.
4900.1700.00).
The following external devices are required for standard operation.
Connect the instrument to the power supply as described in Chapter 3.1.4, "Con-
necting to power", on page 15.
Connect the PC or notebook LAN port to the LAN port of the R&S TSME6 as
described in Chapter 3.1.5.1, "Setting up the LAN connection to the host PC",
on page 17.
Connect the (optional) antenna's SMA connector to the RF IN connector.
NOTE: Do not overload the input power at the RF input connector, otherwise the
input stage could be severely damaged. For maximum allowed values, see the
specifications document.
Connect the GPS antenna to the GPS ANT connector of the instrument for time
synchronization to a GPS signal (3 V, max. 25 mA for active antenna). To ensure
time synchronization of the R&S TSME6, it is required to have a GPS antenna con-
nected.
NOTE: A missing GPS antenna will lead over time to the point that signals cannot
be detected anymore. We recommend connecting a GPS antenna at least once
per month.