Getting started
R&S
®
TSME6
34User Manual 4900.0033.02 ─ 08
● If the firmware version required by the application is not yet available on the
R&S TSME6, the application uploads the required firmware version to the
R&S TSME6 (this step takes about one minute). The application reboots the
R&S TSME6 using the required firmware version, then establishes a connec-
tion to the R&S TSME6.
● If the firmware version required by the application is not yet available on the
R&S TSME6, and other versions occupy all three storage spaces for firmware,
the application must do the following:
a) Clear all storage space for firmware versions on the R&S TSME6 (this step
takes about one minute).
b) Upload the required firmware version to the R&S TSME6 (this step takes about
one minute).
c) Reboot the R&S TSME6 using the required firmware version.
d) Establish a connection to the R&S TSME6.
3.2 Instrument tour
3.2.1 Front panel view
The front panel of the R&S TSME6 does not provide any connectors or control ele-
ments for operation.
Behind the right side of the rear panel (with the ventilation openings), 4 status LEDs
are located. These LEDs display the following states:
●
LEDs ON: R&S TSME6 ready for operation, RF-PLLs initialized correctly
●
LEDs OFF: R&S TSME6 is off or RF-PLLs initialized not correctly
If the fans are off (temperature on the controller board < 60° C), the LEDs are partially
covered by the fan blades.
Instrument tour