Loading...
Vibration-proofed stacking
R&S
®
TSME6
42User Manual 4900.0033.02 ─ 08
5 Vibration-proofed stacking
5.1 Cascading R&S TSMA6/6B and R&S TSME6
To connect R&S TSMA6/6B and R&S TSME6
1. Screw the connecting elements (order no. 4900.0804.00) on the top of the R&S
TSMA6/6B.
Torque: 0.66 Nm ± 0.05 Nm
Secure with liquid plastic
Figure 5-1: Connecting elements
1 = Connecting elements (order no.4900.0804.00)
2. Notice: An insufficient airflow can cause the instrument to overheat, which disturbs
the operation and even cause damage. Make sure that all fan openings are unob-
structed and that the airflow perforations are unimpeded, particularly when you
install the instrument in a rack or packed in a backpack. Keep clear the following
surrounding spaces to the instrument:
Front pane: minimum 2 cm
Left/right panes: minimum 1 cm
Align the connecting elements with the holes on the bottom of an R&S TSME6 and
press the R&S TSME6 down.
Cascading R&S TSMA6/6B and R&S TSME6

Table of Contents