Getting started
R&S
®
TSME6
37User Manual 4900.0033.02 ─ 08
Table 3-4: LAN LED 1 states and their meaning
LED state Description
Green, blinking LAN sending or receiving, or identifying connected device
Table 3-5: LAN LED 2 states and their meaning
LED state Description
Off No connection
Yellow Physical connection established
Yellow, blinking Identifying connected device
DC IN connector
The DC IN connector is required for the DC power supply (10-28 V, max. 1.8 A). For
details, see Chapter 3.1.4, "Connecting to power", on page 15.
RF IN connector (50 Ω)
The optional multi-band RF antenna (700 MHz to 2.6 GHz) or the device providing the
RF signal is connected to the instrument's RF input. The used cable needs an appro-
priate connector (SMA female, 50 Ω input impedance).
Do not overload the allowed maximum input of 20 dBm. Non-compliance destroys the
input mixer.
Pwr./State LEDs
Table 3-6: POWER and STATE LED states and their meaning
STATE LED POWER LED Meaning
Off Off No power supply connected at DC IN
Power supply is off
Power supply < 10 V
Off Yellow Standby
Off Green, blinking
(2 Hz)
FGPA configuration in progress
Red
(up to 5 seconds
during startup)
Green FPGA configuration finished, preparing for start
Off Green R&S TSME6 ready, not connected
Green Green Connected
Green, blinking rap-
idly
Green Measuring
Green, blinking
2 Hz
Green Instrument is identified by the software