52 Operating Instructions arium
®
mini plus | arium
®
mini
Startup
y The dialog box “Fill bag” appears.
t Change the bag (see Chapter 6.5.8, page 52).
t Rinse ultrapure water cartridge (see Chapter 6.5.8, page 52).
6.5.8 Fill
bag
Procedure
t Prepare a container with a sufficient amount of pretreated water (at least 5
liters).
t Place the tank filling tubing into the container and secure it so that its opening is
completely submerged during filling.
t Tap the [w] button.
y The bag is filled with water. Progress is displayed as a percentage.
t To stop the filling process prematurely, for example, if not enough water is
available for filling: Tap the [<] button.
y The dialog box “Fill bag” appears.
t Carry out filling as described above again.
6.5.9
Rinsing the Ultrapure Water Cartridge
During rinsing, the device fills and rinses the ultrapure water cartridge. Air is
removed from the ultrapure water circulation in the process.
y When filling is complete, the dialog box “Preparing to rinse” appears.
t Start the rinsing process again.
t Press the dispense tube into the quick connector of the water outlet.
t Direct the free end of the dispense tube to the drain or into a vessel (at least
6 liters).
y When filling is complete, the dialog box “Start rinsing” appears.
t Tap the [w] button.
1
Fill bag
0DNHVXUHWKHUHLVVXI¿FLHQWZDWHUWR¿OO
WKHEDJ7DS67$57WR¿OOWKHEDJ
w
m
Fill bag
1
Prepare rinsing
&RQQHFWWKHGLVSHQVHWXEHDVGHVFULEHG
LQWKHPDQXDO
OK
11
Start rinsing
7DS67$57WREHJLQWKHULQVLQJ
SURFHVV
w