68 Operating Instructions arium
®
mini plus | arium
®
mini
Cleaning and Maintenance
y The dialog box “Close bag outlet” appears.
t Close the ball cock of the tank outlet tubing.
t Confirm that the tank outlet has been closed with the [
OK] button.
y The arium
®
mini plus automatically begins filling the bag.
The dialog box “Fill bag?” appears on the display of the arium
®
mini.
t To fill the bag of the arium
®
mini:
Tap [YES] button.
Manually filling the bag (only arium
®
mini) (see Chapter 7.3.1, page 59).
t To fill the bag of the arium
®
mini later: Tap the [NO] button.
8.6.2 Removing or Replacing the Final Filter
Requirements
In the “Change consumable” menu, the menu item [
Sterile final filter] is activated.
y The dialog box “Remove filter” appears.
Procedure
t Use the tubing removal tool to push and hold the water outlet quick
connector up.
y The quick connector is unlocked.
t Pull the final filter out of the quick connector.
t Confirm removal of the final filter with the [
OK] button.
Requirements
In the “Change consumable” menu, the menu item [
Sterile final filter] is activated.
The final filter has been removed.
y The dialog box “Connect filter” appears.
Procedure
t To complete the final filter replacement:
Connect final filter (see Chapter 6.4.9, page 42).
Rinse final filter (see Chapter 6.4.10, page 42).
1
Close bag outlet
&ORVHWKHWDQNRXWOHW%DJ2XWOHWDWWKH
UHDURIWKHGHYLFH
OK
1
m
5HPRYH¿OWHU
5HPRYHWKHVWHULOH¿QDO¿OWHU
OK
1
&RQQHFW¿OWHU
&RQQHFWWKHVWHULOH¿QDO¿OWHUDVGHVFULEHG
LQWKHPDQXDO
OK