Cleaning and Maintenance
8.6.4.1 Removing Cartridges
Procedure
t Press the two protruding terminals of the cartridge together and pull the
cartridge forward and out.
t Confirm removal of the cartridge with the [
OK] button.
y If the menu item [UV lamp] is activated in the “Change consumable” menu,
the dialog box “Change UV lamp” appears.
t If required: Change UV lamp (see 8.6.5.2, page 78).
8.6.4.2 Inserting
Cartridges
The arium
®
mini has a pretreatment cartridge and an ultrapure water cartridge.
Procedure
Equipment damage due to impurities!
If dirt or foreign objects enter the ultrapure water circulation, individual device
components can potentially become clogged or wear out faster.
t If possible, do not touch the cartridge connections. This will prevent impurities
from getting into the device.
t Remove the plugs from the two external cartridge connections.
t Slide the cartridge straight into the guide rails of the cartridge holder on the
front side of the device with the corresponding marking (“L” for ultrapure water
cartridge and “R” for pretreatment cartridge). Make sure to hold the cartridge at
the top when doing this.