6.5.11 Rinsing the Final Filter
Requirements
The device is now in dispensing mode.
Procedure
t Place a vessel (at least 6 liters) under the final filter.
t Open the final filter’s vent valve.
t Remove the protective cap from the bell assembly.
t Dispense at least 5 liters of water (the equivalent of a full bag).
This will rinse the final filter.
t If the bag in the arium
®
mini no longer contains a sufficient amount of water or
empties during the rinsing process:
− Wait until the bag has filled to a sufficient level.
− Complete the rinsing process.
t When the rinsing process is complete: Close the vent valve.
t Attach the protective cap to the bell assembly.