78 Operating Instructions arium
®
mini plus | arium
®
mini
Cleaning and Maintenance
8.6.5.2 Inserting the UV Lamp
Touching UV lamps with your bare fingers will cause them to become defective!
Touching the UV lamp with your bare fingers will leave fingerprints. The fingerprints
can become so hot during operation that the UV lamp is destroyed.
t Never touch the glass of the UV lamp with your bare fingers.
t Only hold the UV lamp where it connects to the device or when wearing gloves.
Procedure
t Carefully unpack the new UV lamp without touching the glass with your fingers.
t Insert the new UV lamp all the way into the lamp housing.
t Unscrew the black plastic cover of the UV lamp housing by hand.
t Attach the black connector to the UV lamp. The connector only fits in two
orientations.
t Slide any protruding cable back into the housing.
t Slide the metal retaining clip over the cable and onto the black plastic cover of
the UV lamp housing.
t Insert the closed side of the metal retaining clip into the recess of the black
plastic cover provided.
t Squeeze the open side of the metal retaining clip and insert it into the recesses
provided in the black plastic cover.
y Release the metal retaining clip secures it to the black plastic cover.
t Reconnect the device to the AC power.
y The wizard continues.
t When the dialog box “Insert pretreatment cartridge” appears: Insert cartridge
(see
Chapter 8.6.3.2, page 70).
t When the dialog box “Put the cover back on” appears: Put the front cover on the
device.
t When the dialog box “Prepare rinsing” appears: Rinse ultrapure water cartridge
(see Chapter 6.4.9, page 40).
t When the dialog box “Connect filter” appears: Connect final filter (see
Chapter 6.4.9, page 42).
8.7 Enabling, Disabling or Configuring Reminders for Replacing
Final Filters
If sterile water is constantly required, the final filter must be replaced regularly.
The device can provide a reminder about a pending final filter replacement.
Procedure
t Tap the menu item [
Reminder Endfilter] in the “Care” menu.
y The dialog box “Reminder Endfilter” appears.
t Tap the [Active] button.
y The currently configured replacement interval appears in the column “Reminder
[weeks]”.
t To change the replacement interval: Tap the [t] button.
y The numeric keypad appears.
t Enter replacement interval desired (in weeks). Observe the maximum value of
24 weeks (6 months).
t Confirm the entry with the [l] button.
t To activate the reminder: In the dialog box “Reminder Endfilter”, tap the [l]
button.
>
5HPLQGHU(QG¿OWHU
$FWLYH
,QDFWLYH
5HPLQGHU>ZHHNV@5HPLQGHU(QG¿OWHU