42 Operating Instructions arium
®
mini plus | arium
®
mini
Startup
6.4.9 Connecting the Final Filter
Procedure
t Attach the bell assembly to the final filter.
t Press the final filter into the quick connector of the water outlet.
t Confirm that the final filter has been connected with the [
OK] button.
y The display changes to the dispensing screen.
6.4.10 Rinsing the Final Filter
Requirements
The device is now in dispensing mode.
Procedure
t Place a vessel (at least 6 liters) under the final filter.
t Open the final filter’s vent valve.
t Remove the protective cap from the bell assembly.
t Dispense at least 5 liters of water (the equivalent of a full bag). This will rinse the
final filter.
t If the bag in the arium
®
mini plus no longer contains a sufficient amount of
water or empties during the rinsing process:
Wait until the bag has filled to a sufficient level.
Complete the rinsing process.
t When the rinsing process is complete: Close the vent valve.
t Attach the protective cap to the bell assembly.
y Startup is complete.