76 Operating Instructions arium
®
mini plus | arium
®
mini
Cleaning and Maintenance
Requirements
The filling level is too low for the dispensing volume entered during volume-
controlled dispensing.
y The dialog box “Fill bag?” appears.
Procedure
t Tap the [YES] button.
8.6.5
Replacing the UV Lamp
If a UV lamp is installed in the device, the UV lamp is located behind the
pretreatment cartridge (arium
®
mini plus) or is freely accessible (arium
®
mini).
Requirements
In the “Change consumable” menu, the menu item [
UV lamp] is activated.
y The dialog box “Remove filter” appears.
Procedure
t Remove final filter (see Chapter 8.6.2, page 68).
y The dialog box “Prepare depress.” appears.
t Carry out depressurization (see Chapter 8.8, page 79).
y The dialog box “Remove cover” appears.
t Remove the front cover.
arium® mini plus
y The dialog box “Remove pretreatment cartridge” appears.
t Remove cartridge (see Chapter 8.6.3.1, page 69).
y The dialog box “Change UV lamp” appears.
t Remove UV lamp (see Chapter 8.6.5.1, page 77).
t Insert UV lamp (see Chapter 8.6.5.2, page 78).
arium® mini
y The dialog box “Change UV lamp” appears.
t Remove UV lamp (see Chapter 8.6.5.1, page 77).
t Insert UV lamp (see Chapter 8.6.5.2, page 78).
1
Fill bag?
7KHEDJFRQWDLQVORIZDWHU:RXOG\RX
OLNHWR¿OOLWQRZ"
NO YES
1
Change UV lamp
3OHDVHGLVFRQQHFWWKHGHYLFHIURPWKH
SRZHUVXSSO\&KDQJHWKH89ODPSDV
GHVFULEHGLQWKHPDQXDO