60 Operating Instructions arium
®
mini plus | arium
®
mini
Operation
Requirements
The bag contains water.
Bag filling must be initiated using the menu.
Procedure
t Tap the
symbol [Fill bag].
y The dialog box “Fill bag” (Step 1) appears.
t Check whether the tank filling tubing is connected to the “Inlet” connection.
If required: Connect the tank filling tubing.
ATTENTION! Insufficient water quality possible due to contaminated tank filling
tubing!
t Check the tank filling tubing for contamination. If required:
Change the tank filling tubing.
Change the bag (see Chapter 8.6.1, page 66).
t Tap the [OK] button.
y The dialog box “Fill bag” (Step 2) appears.
t Prepare a vessel with pretreated water (at least 5 liters).
t Place the tank filling tubing into the container and secure it so that its opening is
completely submerged during filling.
t Tap the [w] button.
y The bag is filled with water. Progress is displayed as a percentage.
t To stop the filling process prematurely, for example, if not enough water is
available for filling: Tap the [>] button.
y The dialog box “Fill bag” (Step 2) appears.
t Carry out filling as described above again.
y When filling is complete, the dialog box “Filling finished” appears.
t Verify that the free end of the tank filling tubing is secure. If required: Remove
the tank filling tubing from the “Inlet” connection at the rear of the device and
keep it in a safe place.
t Confirm the bag filling process with the [OK] button.
y The display changes to the dispensing screen.
7.3.2 Activating or Deactivating Standby Mode
If no water is dispensed for a long time, the device automatically switches to standby
mode. This ensures more economical and ecological operation. In standby mode, the
pre-stage is no longer active and ultrapure water is no longer in circulation. Standby
mode can also be activated manually.
j
i
k
'LVSHQVH &DUH
)LOOEDJ
1
m
3UHSDUH¿OOLQJ
&RQQHFWWKHWDQN¿OOLQJWXEH,QOHWRQ
WKHUHDURIWKHGHYLFHDVGHVFULEHGLQ
WKHPDQXDO
OK
1
m
Fill bag
(QVXUHWKDW\RXKDYHHQRXJKZDWHUWR
¿OOWKHEDJ7DS³67$57´WR¿OOWKHEDJ
w
m
Fill bag

1
)LOOLQJ¿QLVKHG
5HPRYHWKHWDQN¿OOLQJWXEH,QOHWRQWKH
UHDURIWKHGHYLFH
OK

Table of Contents

Related product manuals