80 Operating Instructions arium
®
mini plus | arium
®
mini
Cleaning and Maintenance
8.9 Carrying Out Venting
During venting, the device fills and rinses the ultrapure water cartridge.
Air is removed from the ultrapure water circulation in the process.
The ultrapure water circulation must be rinsed under the following conditions:
− when the water quality displayed during operation fluctuates constantly
− if the device was taken out of service for an extended period
Procedure
t Tap the menu item [
Venting] in the “Care” menu.
y The dialog box “Remove filter” appears.
t Remove final filter (see Chapter 8.6.2, page 68).
t Confirm removal of the final filter with the [OK] button.
y The dialog box “Prepare rinsing” appears.
t Place a vessel (at least 1 liter) under the water outlet. Alternatively, connect the
dispense tube and run the free end to the drain.
t Tap the [OK] button.
arium
®
mini plus
y If the bag in the arium
®
mini plus no longer contains a sufficient amount of
water or empties during the rinsing process, the bag is filled automatically.
y When filling is complete, the dialog box “Prepare rinsing” appears.
t Start the rinsing process again.
arium
®
mini
y If the bag in the arium
®
mini no longer contains a sufficient amount of water or
empties during the rinsing process, the dialog box “Fill bag” appears.
t Manually filling the bag (only arium
®
mini) (see Chapter 7.3.1, page 59).
y After the rinsing process starts, the remaining rinse time is displayed in minutes.
t In order to cancel the rinsing process, e.g. if a sufficiently large vessel is
unavailable: Tap the [<] button.
y The dialog box “Prepare rinsing” appears again.
t Start the rinsing process as described above again.
1
5HPRYH¿OWHU
5HPRYHWKHVWHULOH¿QDO¿OWHU
OK
1
Prepare rinsing
3ODFHDVXI¿FLHQWO\ODUJHFRQWDLQHUOLWHU
XQGHUWKHZDWHURXWOHW
OK
1
l
3UHSDULQJWR¿OOWKH
bag
:DLWXQWLOWKHEDJKDVEHHQVXI¿FLHQWO\
¿OOHG5LQVLQJFRQWLQXHVDXWRPDWLFDOO\
m
Fill bag
1
Fill bag
0DNHVXUHWKHUHLVVXI¿FLHQWZDWHUWR
¿OOWKHEDJ7DS³67$57´WREHJLQ¿OOLQJ
WKHEDJ
w
m
Rinsing
PLQXWHV