64 Operating Instructions arium
®
mini plus | arium
®
mini
Cleaning and Maintenance
8.3 Maintenance Schedule
Depending on the volume of water dispensed, it may be necessary to change the
consumables more often than specified in the maintenance schedule. If, for example,
sterile water is required, the final filter must be replaced regularly.
Tip
It is advisable to replace different consumables in the same maintenance step.
This saves time and water.
t Sensibly plan consumables replacement.
Interval Consumables Item
6 months Bag Replacing the bag
(see Chapter 8.6.1, page 66)
1 - 24 weeks
(depending on
the application)
Final filter Replacing the final filter
(see Chapter 8.6.2, page 68)
6 months Pretreatment cartridge
(only arium
®
mini plus)
Replacing the pretreatment
cartridge (see Chapter 8.6.3,
page 69)
Max. 6 months
(depending on the
volume of water
dispensed)
Ultrapure water cartridge Replacing the ultrapure water
cartridge (see Chapter 8.6.3,
page 69)
12 months UV lamp (optional) Replacing the UV lamp
(see Chapter 8.6.5, page 76)
8.4 Opening the “Care” Menu
The “Care” menu contains all the steps for the care and maintenance work.
Procedure
t Tap the
symbol [Care].
y The “Care” menu appears.
t Displaying reminders (see Chapter 8.5, page 65).
t Replacing consumables (see Chapter 8.6, page 65).
t Enabling, disabling or configuring reminders for replacing final filters (see
Chapter 8.7, page 78).
t Manually carrying out depressurization (see Chapter 8.8, page 79).
t Carrying out venting (see Chapter 8.9, page 80).
j