74 Operating Instructions arium
®
mini plus | arium
®
mini
Cleaning and Maintenance
t Press the cartridge firmly into the cartridge holder until it locks into the guide
rails with a distinct clicking sound. Check that the cartridge has been installed
securely by turning it slightly.
t Confirm the insertion of the cartridge with the [
OK] button.
Requirements
In the “Change consumable” menu on the arium
®
mini, the menu item [Ultrapure
water cartridge
] is activated.
y The dialog box “Change cartridge” appears and asks you to replace the second
cartridge.
Requirements
Cartridge replacement is complete.
In the “Change consumable” menu, the menu item [
UV lamp] is activated.
y The dialog box “Change UV lamp” appears.
Procedure
t If required: Change UV lamp
(see Chapter 8.6.4, page 72).
Requirements
Cartridge replacement is complete.
In the “Change consumable” menu, the menu item [
UV lamp] is not activated.
y The dialog box “Put the cover back on” appears.
Procedure
t Put the front cover on the device.
t Confirm that the cover has been put on with the [
OK] button.
y The dialog box “Prepare rinsing” appears.
t Prepare rinsing: Connect the dispense tube (see Chapter 6.4.9, page 42)
1
Put the cover back on
3OHDVHSXWWKHIURQWFRYHUEDFNRQ
OK
1
Prepare rinsing
$WWDFKWKHGLVSHQVHWXEHDVGHVFULEHGLQ
WKHPDQXDO
OK