EasyManua.ls Logo

Sartorius arium mini - Removing the UV Lamp

Sartorius arium mini
97 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating Instructions arium
®
mini plus | arium
®
mini 71
Cleaning and Maintenance
Requirements
In the “Change consumable” menu of the arium
®
mini, the menu item
[Ultrapure water cartridge] is enabled.
y The dialog box “Change cartridge” appears and asks you to replace the second
cartridge.
Requirements
Cartridge replacement is complete.
In the “Change consumable” menu, the menu item [
UV lamp] is activated.
y The dialog box “Replace UV lamp” appears.
Procedure
t If required: Replace the UV lamp
(see Chapter
8.6.5, page 76).
Requirements
Cartridge replacement is complete.
In the “Change consumable” menu, the menu item [
UV lamp] is not activated.
y The dialog box “Put the cover back on” appears.
Procedure
t Attach the front cover to the device.
t Confirm that the cover has been attached with the [
OK] button.
y The dialog box “Prepare rinsing” appears.
y Prepare rinsing: Connect sampling tubing (see Chapter 6.4.9, page 42)
t Open the ball cock.
t Start the pre-filter cartridge rinsing process.
t When the rinsing process is complete:
Close the ball cock.
Start automatic filling of the bag.
y When the filling process is complete, the dialog box „Start rinsing“ appears.
t Rinse ultrapure water cartridge (see Chapter 6.4.9, page 42).
t Close the tank outlet (Bag Outlet) on the rear side of the device
1
Put the cover back on
3OHDVHSXWWKHIURQWFRYHUEDFNRQ
OK
1
Prepare rinsing
3ODFHDVXLWDEOHYHVVHOOXQGHUWKH
GLVSHQVLQJXQLW
w
1
Start rinsing
7RVWDUWWKHULQVLQJSURFHVVWDSWKH³6WDUW´
EXWWRQ
START
1
Close bag outlet
&ORVHWKHWDQNRXWOHW%DJ2XWOHWRQWKH
UHDUVLGHRIWKHGHYLFH
START

Table of Contents