21
SARTORIUS Basic 
plus
 
 Semimicro-Balances
bp-a14a.tif
bp-a-19a.tif
bp-a-16a.tif
BP210D
40
80
210
g
0,01
0,01
mg 20 g 0,02 mg 200 g 0,4 mg E2
BP300S 300
200
200
E2
BP210d_a.xls
Internal Span Adjustment
– Make sure that the menu code  »1 9 3 - internal calibration/adjustment
accessible « is set (see page 27, “Balance Menu Code Settings)”.
– Press the t key to tare the balance; then press q to activate the
calibration/adjustment function.
– The built-in, motorized calibration/adjustment weight is applied, then
removed automatically; the weight is stored; and the balance then returns
to the standard weighing mode.
– After completing the adjustment procedure, reset the » 9--1 « code in the
balance operating menu.
– Slide the menu access switch back to the left.
Replace the capped plug to close the opening.
Table of Adjustment Data
Only valid when the measuring regulations stipulated in the “Standard Operating Procedure” WKD-038 are maintained.
The following applies to these specifications: WKD-038 and Test Record WKD-038-1/09.96
BP210d_a.xls
Span accuracy Linearity TCS
Model
Adjustment
weight
Test
weight
Permissible
tolerances
Tare
weight
Test
weight
Permissible
tolerances
Reproducility Off-center load
Model Weighing
capacity
Readability
Test
weight
Permissible
tolerances
Test
weight
Permissible
tolerances
BP210D internal 200 g 0,4 mg
10/20
/30
50/10
g
10
50
g
0,03
0,2
mg 1
BP300S internal 300
70/15
70
1