Operating Instructions Cubis
®
MCA 107
Technical Data
15.2.4 Electromagnetic Compatibility
Interference resistance
Suitable for use in industrial areas
Transient emissions
Class B
Suitable for use in residential areas and areas that are connected
toa low voltage network that also supplies residential buildings.
15.3 Ambient Conditions
15.3.1 Installation Site
Unit Value
Installation site
Standard laboratory rooms
Installation site according to IEC 60259-1,
maximumaltitudeabove sea level
m 3000
For indoor use only
Temperature
In operation °C +5–+40
In operation for conformity-assessed devices:
See information on the device’s ID plate
During storage and transport °C -20–+60
Relative humidity
At temperatures of up to 31°C % 80
Then linear decrease from 80% at 31°C to 50% at 40°C
No heat from heating systems or direct sunlight
No drafts from open windows, AC systems, or doors
No vibrations
No “heavy traffic” areas (personnel)
No electromagnetic fields
No dry air