Operating Instructions Cubis
®
MCA 63
System Settings
7.7.4 Parameters in the “Weighing and Print Profiles / Weighing” Menu
Parameter Setting Values Explanation
Ambient
conditions
Very stable Sets the ambient conditions to “very stable”: Activates a fast
change in the weight values in the event of a load change with
ahigh output rate.
Recommended for the following work environment:
Very stable table near the wall
Closed and calm room
Stable Sets the ambient conditions to “stable”.
Recommended for the following work environment:
Stable table
Slight movement in the room
Slight draft
Unstable* Sets the ambient conditions to “unstable”: Activates the delayed
change in weight values with a reduced output rate.
Recommended for the following work environment:
Simple office desk
Room with moving machinery or personnel
Slight air movement
Very unstable Sets the ambient conditions to “very unstable”: Activates a
significantly delayed change in the weight values and long wait
for stability with a further reduction in the output rate.
Recommended for the following work environment:
Noticeable and slow floor vibrations
Noticeable building vibrations
Weighed goods moved
Very strong air movements
Application filter Weighing
(final readout)*
Activates a filter that enables a fast change in the display for very
fast load changes.
Display changes with minimal load changes (in the digit range)
occur more slowly.
Dosing (filling mode) Activates a filter that enables a very fast change in the display
with minimal load changes (e.g. when dosing or filling vessels).
Sensor mode (low filter) Activates a weak but fast filter that always behaves in the same
way for load changes (e.g. when dosing in automated systems).
Dynamic mode
(withoutfilter)
Deactivates the active application filter.
*Factory setting for the “Standard Weighing” factory-set weighing profile

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis MCA225P Series and is the answer not in the manual?

Sartorius Cubis MCA225P Series Specifications

General IconGeneral
BrandSartorius
ModelCubis MCA225P Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals