112 Operating Instructions Cubis
®
MCA
Technical Data
15.7.3 Models MCA224S | MCA124S | MCA5203S | MCA5203P
MCA224S MCA124S MCA5203S MCA5203P
Unit Value Value Value Value
Scale interval (d) mg 0.1 0.1 1 1/2/5
Maximum capacity (Max) g 220 120 5200 1200/2400/
5200
Repeatability at 5% load
Standard deviation of the load values,
tolerance
mg 0.07 0.1 1 1
Standard deviation of the load values,
typical value
mg 0.04 0.04 0.5 0.5
Repeatability at approx. maximum capacity
Standard deviation of the load values,
tolerance
mg 0.07 0.1 1 1
Standard deviation of the load values,
typical value
mg 0.05 0.05 0.6 0.6
Linearity deviation
Tolerance mg 0.2 0.2 5 5
Typical value mg 0.13 0.13 2 3
Deviation when load is off-center,
positions according to OIML R76
Test weight g 100 50 2000 2000
Tolerance mg 0.2 0.2 2 2
Typical value mg 0.12 0.12 1 1
Sensitivity drift between +10°C and +30°C ppK 1 1 1 1
Tare maximum capacity:
Less than 100% ofmaximum capacity
Accuracy class according
toDirective2014/31/EU
I I I I
Verification scale interval (e) according
toDirective 2014/31/EU
mg 1 1 10 10
Minimum load (Min) according
toDirective2014/31/EU
mg 10 10 100 100
Minimum initial weighing according to USP
(United States Pharmacopeia), Chap. 41
Optimum minimum initial weighing mg 82 82 820 820
Typical minimum initial weighing mg 82 82 1000 1000
Typical stabilization time s 1 1 1 1
Typical measurement time s 3 3 2 2