Operating Instructions Cubis
®
MCA 81
System Settings
7.7.25 Parameters in the “Device Settings / Electronic Signature” Menu
Parameter Setting Values Explanation
Sign report Off* If the QAPP extension has been activated:
Deactivates electronic signature.
On If the QAPP extension has been activated:
Activates electronic signature.
* Factory setting
7.7.26 Parameters in the “Device Settings / Draft Shield” Menu
(Only for Devices with a Motorized Draft Shield)
Parameter Setting Values Explanation
Left/right palm key Off, no function Deactivates the function for both palm keys.
Same function Both palm keys perform the same learned draft shield function.
The two palm keys cannot be assigned functions independently
of each other.
Separate functions* The two palm keys perform independent learned draft shield
functions. The two palm keys can be assigned functions
independently of each other.
Automatic mode Off, no function* Deactivates the automatic function.
Closing ->
Function->open
Closes the draft shield, carries out the function, and then opens
the draft shield again.
Close->Function Closes the draft shield and then carries out the function.
* Factory setting
7.7.27 Parameters in the “Device Settings / Ionizer” Menu (Only for Devices with an Ionizer)
Parameter Setting Values Explanation
Activation ionizer Off, no function* Deactivates the ionizer.
Manual on with key The [Ionizer] button starts an ionization process.
Autom. on after closing
draft shield
If the draft shield is closed: The ionization process starts
automatically.
Ionizer intensity Weak Sets the intensity of the ionization process to “weak”.
Medium* Sets the intensity of the ionization process to “medium”.
Strong Sets the intensity of the ionization process to “strong”.
Start-up duration Input: 1 to 60 seconds Sets the duration of the ionization process (factory setting:
8seconds).
* Factory setting

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis MCA225P Series and is the answer not in the manual?

Sartorius Cubis MCA225P Series Specifications

General IconGeneral
BrandSartorius
ModelCubis MCA225P Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals