System Settings (Setup Menu)
Operating Instructions Moisture Analyzer MA160
  Device Information
  The manufacturer, model, serial number, and software version of the moisture 
analyzer are displayed here.
 Calibration/Adjustment
 
Calibration/Adjustment
MA performance test
Weighing system adjustment
Calibration report
Cal./Adj. the Heating Module
  The settings for calibration and adjustment of the moisture analyzer can be defined 
here.
Under MA Performance Test:
Test functions for different temperature ranges are available here. You can only use 
these functions with the ReproEasy pad (accessory).
Under Weighing system adjustment:
t  Select the option of your choice.
–  CAL-Extern: Carry out calibration with an external calibration weight  
(see section “Calibrating the Weighing System” in chapter “Calibration and 
Adjustment”).
–  Weighing only: Use the weighing function to use the moisture analyzer as 
a simple lab balance (see section “Weighing” in chapter “Calibration/
Adjustment”).
Under Heater module cal./adj.:
You can calibrate and adjust the heating module here in connection with the  
YTM15MA heating adjustment set. (see section “Calibrating/Adjusting the Heating 
Module” in chapter “Calibration and Adjustment”).
Under Calibration report:
You can view the current calibration reports on the MA performance tests, weighing 
system adjustment, and heating module calibration and adjustment. Up to 99 logs 
per day are saved. Calibration reports older than 30 days are deleted. Only the cur-
rent calibration report is saved for each calibration or adjustment of the heater.
Device information
Manufacturer:  Sartorius
Model:  MA160
Serial number:  0000021087
Version BAC:  00-53-01.01.00
Version APC:  01-80-01.09.01
Version MCU:  07-40-05.00.00
Calibration report
MA performance test
Weighing system adjustment
Heater calibration
Heater adjustment