40
Moisture Analysis
Operating Instructions Moisture Analyzer MA160
+ 4.000 g
Place pan and two filters, close
hood for taring
t Place an empty sample pan on the analyzer, along with one or two filters if
necessary. The number of filters depends on how many filters are set in the
parameters for the current method.
t Close the hood.
y The moisture analyzer tares automatically.
Open hood
0.000 g
t Open the hood.
Fill in sample and close hood to
start
- 0.000 g
t Place the sample on the sample pan or filter(s).
Fill in sample
+ 1.929 g
2.0 g
y If you have activated the "Initial weight" option in the method parameters
(setting "On" or "Mandatory"), a target value display appears.
t Add the sample to the sample pan slowly. When the sample quantity reaches
the preset range, the target value display turns green.
If the “Mandatory” setting has been selected, the measurement only starts if the
initial weight is within the permissible range.
If the “On” setting has been selected, the measurement also starts if the initial
weight is outside of the permissible range.
Close hood to start
+ 1.929 g
2.0 g
t Close the hood.
y The moisture analysis starts automatically.
During the measurement, the process status light on the analyzer flashes, if the
light has not been switched off in the menu.