EasyManua.ls Logo

Sartorius MA160 - Printing Out the Analysis Report; Removing a Sample

Sartorius MA160
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
42
Moisture Analysis
Operating Instructions Moisture Analyzer MA160
Printing Out the Analysis Report
If a lab printer is connected, you can print out the result of the moisture analysis.
a
t Select a.
y The analysis report is printed out.
The settings for the analysis report can be configured in the system settings [see
section “Printout” in chapter “System Settings (Setup Menu)”]. Further information
and an example of an analysis report can be found in chapter “ISO/GLP Report.”
Removing a Sample
Danger of Burns due to Hot Sample Pan!
The sample and sample pan may be very hot.
Do not touch the sample pan.
Use sample forceps (included in delivery) to remove the sample.
Remove sample
Remove old sample from sample
chamber.
y When you have exited the measurement by selecting OK, you can remove the
sample from the moisture analyzer.
t Open the hood.
t Remove the sample pan from the moisture analyzer using sample forceps.
t Close the hood.
t Allow the sample to cool down and dispose of it properly.

Related product manuals