46
Moisture Analysis
Operating Instructions Moisture Analyzer MA160
Sample type
Specify sample type
;
y In the last step, enter the sample type.
t Select t.
Sample type
organic
inorganic
unknown
t Select the desired option:
– organic
– inorganic
– unknown
t Select l to confirm.
Test the method
Method successfully created. Would you
like to test the method now?
Temperature: 130 °C
NOYES
y The method can then be tested.
t Select the desired option:
To test the method with the suggested temperature: Select YES. Moisture
analysis is then carried out with the set parameters (see “Performing the
Moisture Analysis” on page 39).
To save the method without testing: Select NO and then o to return to the
method menu.
+
Standard
Coffee creamer
100 °C
Fully automatic
Oats
80 °C
Wood glue
110 °C
Fully automatic
130 °C
Fully automatic
Ketchup
110 °C
Fully automatic
y The newly defined method now appears in the method menu. It is not available
yet and therefore marked with a H symbol.
Creating a New Method Manually
To define each parameter for a method yourself, create the method manually:
t Select + in the method menu.
New method
Manual input
Assistant
Import
t Select Manual input.