47
Moisture Analysis
Operating Instructions Moisture Analyzer MA160
Method name
Enter name for new method
y First enter a name for the new method.
t Select t.
qwer tzuiop
asdfgh
yxcvbnm- /
123
Ke
t Enter the desired name.
t Select l to confirm.
New method
Heating program
End
Initial weight
Standard drying
110 °C
Fully automatic
Off
Method name Ketchup
y The new method is created and displayed with the default parameters.
New method
Heating program
End
Initial weight
Standard drying
110 °C
Fully automatic
Off
Result display
Moisture in %M
Method name Ketchup
t Select the parameters you wish to change and configure the desired settings.
t Scroll through the list if necessary and change further parameters.
t Select l to confirm.
Method name: Ketchup
Type: Standard drying, 110 °C
End: Fully automatic
Initial weight: Off
Result display: Moisture in %M
Target value: Off
Save new method
NOYES
y The parameters of the new method are displayed.
You can scroll through the list to view all parameters.
t Select the desired option:
– To change further parameters: Select <. This takes you back to the list of
parameters.
– To discard the new method: Select NO.
– To save the new method: Select YES.