49
Moisture Analysis
Operating Instructions Moisture Analyzer MA160
New method
Heating program
End
Initial weight
Standard drying
85 °C
Fully automatic
Off
Method name Coffee
y The parameters of the method are displayed.
If the moisture analyzer’s memory already contains a method with the same
name, a message appears, indicating that the imported method must be
renamed.
t To accept the imported method, select l.
Save new method
Method name: Coffee
Type: Standard drying, 85 °C
End: Fully automatic
Initial weight: Off
Result display: Moisture in %M
Target value: Off
NOYES
y The parameters of the new method are displayed.
t To save the method, select YES.
y The imported method now appears in the method menu. It is not available yet
and therefore marked with an H symbol.
Testing Methods
Evaluation
Result is within the tolerance range
Result: 85.15 %M
Target value: 85.00 %M
Tolerance range: 84.50 %M - 85.50 %M
Are you satisfied with this result?
NOYES
y If the target value and initial weight options are activated, the measurement/
method is evaluated and the result of the evaluation is displayed.
t Select the desired option:
If you are satisfied with the result, select YES. You can then release the
method.
If you are not satisfied with the result: Select NO and run the test again.
Test the method
Method successfully created. Would you
like to test the method now?
Temperature: 130 °C
NOYES
t Select the desired option:
To test the method with the suggested temperature: Select YES. Moisture
analysis is then carried out with the set parameters (see “Performing the
Moisture Analysis” on page 39).
To save the method without testing: Select NO and then o to return to the
method menu.