  Parameter for saving weight values
  The reference weight is saved when the 
scale has stabilized.  
“Stability” is defined as the point at 
which the fluctuation of a measured 
value lies within a defined tolerance 
range. The narrower the tolerance 
range, the more stable the platform 
is at “stability”.
  In Setup, under:
Appl
: Z nM: 3.11.  
Signum
®
 3: 
appl1
:  
Neutr.M:3.11. 
You can define whether the value is 
saved when “standard stability” is 
reached, or only at “increased stability” 
(narrower tolerance range). If you 
select “increased stability” the reference 
weight saved will be more accurate and 
the results more reproducible, but the 
response time of the weighing platform 
might be longer.
  Accuracy Level for Calculation of 
Reference Value
  The resolution applied for calculating 
the reference weight is defined in 
Setup under:
Appl
: Z nM: 3.9.  
Signum
®
 3: 
appl1
:  
Neutr.M: 3.9. 
The resolution for calculating the 
reference weight is increased if 
“+1  decimal place,” “+2 decimal 
places” or “Internal resolution” is 
selected. With the “+1 decimal place” 
setting, the net value is determined 
to one additional decimal place (i.e., 
display accuracy x 10); “+2 decimal 
places” increases display accuracy x 
100, and “Internal resolution” uses 
the maximum resolution available. 
  Decimal Places in Displayed Result
  In neutral measurement, not only whole 
numbers but also decimal numbers 
(for example, 1.25 o electrical cabling) 
can be displayed. The number of 
decimal places displayed in neutral 
measurement is configured in Setup 
under:
Appl
: Z nM: 3.10. 
Signum
®
 3: 
appl1
:  
Neutr.M: 3.10.
  Minimum load
  The minimum load required for 
initialization of the weighing platform 
is configured in Setup under:
Appl: Z nM: 3.6.  
Signum
®
 3: 
appl1
:  
Neutr.M: 3.6. 
Once the limit is exceeded by the load, 
initialization can begin.  
If the load on platform is too light, 
the following will occur when you try 
to save a value:
–  The error code Inf 29 is displayed
–  A warning signal is emitted (double-
beep)
–  The weighing platform is not initialized
–  The preset calculation factor is saved
  The minimum load required for 
automatic taring of the container 
weight on the platform (first weight) 
is configured in Setup under:
  Appl
: Z nM: 3.5.  
Signum
®
 3: 
appl
: m.weigh: 3.5.
  You can choose from the following 
10 levels for this setting:
  1 digit
  2 digits
  5 digits
  10 digits
  20 digits
  50 digits
  100 digits
  200 digits
  500 digits
  1000 digits
  The “digits” here refer to the scale 
intervals in the connected weighing 
platform. If the interval of the 
connected platform is 1 g, for example, 
and 1000 digits are required, you must 
place at least 1000 g (=  1000 intervals 
=1000 digits) on the weighing platform 
for initialization.
  Neutral Measurement with 
Two Weighing Platforms
  You can use two weighing platforms 
simultaneously with the Neutral 
Measurement application. When using 
two platforms, you can choose from 
the following operating modes:
–  Neutral measurement with two 
weighing platforms
–  Neutral measurement with one 
reference platform and one weighing 
platform
  Neutral measurement with two 
platforms of the same type:
  Use this operating mode to measure 
different types of sample material with 
different weights. For example, measure 
the lighter-weight samples on one 
platform and the heavier samples on 
another.
  You can define one of the two 
platforms as the default scale.  
This is configured in Setup, under:
  setup: 
utilit: 
8.11. : (main scale)
  This is the first platform active when 
you switch on the Signum, regardless of 
the setting for automatic initialization 
of the Neutral Measurement 
application.
  Neutral Measurement with One 
Reference Platform and One Weighing 
Platform: 
  In this operating mode, the reference 
platform is a high-resolution weighing 
platform with a relatively low maximum 
capacity. The other platform is used 
for weighing heavier samples, and has 
a high capacity with a relatively low 
resolution.
  This allows you to both determine the 
reference weight with high resolution; 
i.e., very precisely, and to measure large 
samples, without requiring an expensive 
high-resolution, high-capacity weighing 
platform.
  The system can be configured to 
switch automatically to the reference 
platform for initialization. Following 
initialization, the platform for larger 
amounts is automatically activated. 
  The definition of one weighing platform 
as a reference platform is configured in 
Setup, under:
Appl
: Z nM: 3.13. 
Signum
®
 3: 
appl1
:  
Neutr.M: 3.13.