58
Example:
Measuring 25 m of electrical cable.
Settings (different from the factory settings):  
Setup: Application 1: Neutral measurement
Setup: Printout; PRTPROT 7.6. 
then select the menu line items of your choice
  1.)  Place empty container on the 
platform
)
  2.)  Tare the scale  
Note: If the automatic tare 
function is enabled, you do 
not need to press the ) key 
to tare the platform; the tare 
weight is saved automatically 
when you place the container 
on the platform
248
  3.)  Signum
®
 3 only: 
Enter the weight of 1 meter of 
cable using the keypad  
(in this example, 248 g)
O
  4.)  Save the value entered as the 
reference weight
  5.)  Place the desired amount of 
cable in the container
    Read the result
p
  6.) Print the results
nRef  +        1 o  Configured printout: 
wRef  +    0.248 kg  see page 81
G#    +    6.794 kg
T     +    0.541 kg
N     +    6.253 kg
Qnt           25 o 
--------------------