Appendix III: Command Summary
105
CATEGORY COMMAND
DESCRIPTION
GetSamples:b,e
or DDb,e
Upload data from scan b to scan e.
GetCast:x or
DCx
Profiling mode only. Upload data from cast x.
Data Upload
Stop logging
before
uploading.
GetHeaders:b,e
or DHb,e
Upload header b to header e.
SL
Output last sample from buffer and leave power on.
SLT
Output last sample from buffer, then take new sample and store
data in buffer. Leave power on.
TS
Take sample, store data in buffer, output data, and leave power
on.
TSS
Take sample, store in buffer and FLASH memory, output
data, and turn power off.
TSSOn
Take sample, store in buffer and FLASH memory, output
data, and leave power on.
Polled
Sampling
GetLastSamples:
x
Output last x samples from FLASH memory.
TC
Measure conductivity, output converted data.
TCR
Measure conductivity, output raw data.
TT
Measure temperature, output converted data.
TTR
Measure temperature, output raw data
TP
Measure pressure, output converted data.
TPR
Measure pressure, output raw data.
TV
Measure 6 external voltage channels, output converted data.
TVR
Measure 6 external voltage channels, main battery voltage,
back-up lithium battery voltage, external current, pressure
temperature; output raw data.
TF
Measure frequency (optional Quartz pressure),
output converted data.
TFR
Measure frequency (optional Quartz pressure), output raw data.
T38
Measure SBE 38, output converted data.
PumpOn
Turn pump on for testing purposes.
Testing
Takes and
outputs
100 samples
for each test.
Press Esc key
(or send break
character) to
stop test.
PumpOff
Turn pump off for testing purposes.
Note:
Use the Upload menu to
upload data that will be
processed by SBE Data
Processing. Manually
entering the data upload
command does not
produce data with the
required header
information for processing
by SBE Data Processing.
112