METIS M308 / M309 / M313 / M316 / M318 / M323 / H309 / H316 / H318 (12-pin)
Mechanical Installation
20
5.6.2 Manually Adjustable or Motorized Focus Optics with camera module
Model
Temperature ranges
[°C]
M308: all tempera-
ture ranges
M318: 100-700°C
H316: ZSC <400°C
H318: 120–520°C
M309: all ranges
M316: all ranges
M318: 150–1200°C
180–1300°C
H309: all ranges
H316: ZSC ≥400°C
H318: 180–800°C
Spot size diameter M [mm]
M308 / M309:
OV09-D1
M316 / M318:
OV09-D2
ZSC = Zero-scale temperature
Aperture diameter D: Full scale temperature value ≤ 1400 °C: 14–16 mm; > 1400 °C: 7–8 mm
(D is dependent on the optics pull-out: optics pulled out: higher value, optics inserted: smaller value)
5.6.3 Manually Adjustable or Motorized Focus Optics Model M313
Model
Temperature ranges
[°C]
M313: 500–2600°C
550–3000°C
600–3500°C
M313: 450–1800°C
500–2200°C
Spot size diameter M [mm]
Aperture diameter D: Full scale temperature value ≤ 2200 °C: 14–16 mm; > 2200 °C: 7–8 mm
(D is dependent on the optics pull-out: optics pulled out: higher value, optics inserted: smaller value)