Maintenance
for the LuMon System
75 of 93
To clean the LuMon Monitor and the belt connector proceed as follows:
Gently wipe the surfaces of the monitor and belt connector and cable using the listed agents with a
soft cloth, and do not allow excessive fluid or moisture to enter the monitor or connector ports.
Surfaces should be cleaned prior to disinfection.
Allow surfaces to dry fully before placing cleaned items into service or prior repackaging when
shipping them (11.5).
A. Surface cleaning & low level disinfection
Mikrozid (Schülke Mayr) pre-soaked
wipes
25% Ethanol, 35% Isopropanol
Carefully wipe the outer surfaces with alcohol
wipes and allow to dry.
WEBCOL Alcohol Prep Pads
(Kendall)
70% Isopropanol
B. Low level disinfection
Terralin Protect (Schülke&Mayr)
22 % quaternary ammonium
compounds (Benzyl-C12-16-
alkyldimethyl-, Chloride),
17 % 2-Phenoxyethanol,
0,9 % Aminoalkylglycine,
5 15% non-ionic tensides.
Prepare a 2% solution (e.g. 980 ml deionized water
plus 20 ml Terralin Protect). Carefully wipe the
outer surfaces. Wait 15 min. Perform a final wipe
using using 70% Isopropanol or presoaked
Mikrozid wipes. Allow to dry.
WARNING
Before cleaning the LuMon Monitor, always switch it off and disconnect it from AC mains power.
WARNING
Do not expose the LuMon Monitor to heavy moisture and do not allow any fluids to enter the LuMon
Monitor. Plugs and connectors have to be kept meticulously clean and dry at all times. Failure to do so increases
the risk of electrical shock and/or may result in damage to the device.
CAUTION
Use of other cleaning and disinfection agents than recommended may cause damage and/or deterioration of
the device erials and device failure may result.
CAUTION
Applying excessive mechanical force to the device during cleaning ials and
device failure may result.
CAUTION
Do not touch, press or rub the surfaces of the LuMon Monitor or belt connector with abrasive cleaning
compounds, instruments, brushes, rough surface materials, or bring them into contact with any that could
scratch the surfaces of the LuMon Monitor or belt connector.
CAUTION
Do not use petroleum-based or acetone solutions, or other harsh solvents, to clean the LuMon Monitor or belt
connector. These substances can attack the device materials and device failure may result.
CAUTION
The LuMon System and its accessories are provided non-sterile. Do not sterilize any parts of the equipment by
irradiation, steam or ethylene oxide. Do not autoclave or pressure sterilize.
NOTE
The LuMon Monitor screen supports finger touch operation only. The use of rigid or sharp instruments on the
touch screen can permanently damage the display.
Name brand/ manufacturer
Composition
Recommended procedure
Name brand/ manufacturer
Composition
Recommended procedure