NOTE
Belts are specified for single-patient use only and must be discarded after use according to local regulations for
medical waste.
11.3 Replacing the fuses of the LuMon Monitor
Before replacing the fuses of the LuMon Monitor, end monitoring (10.3), switch the LuMon Monitor off and
disconnect the LuMon Monitor from AC mains power.
To replace the fuses, open the fuse holder (Figure 5-2) with a screwdriver and replace the fuses with specified
fuses only (14.1).
11.4 Disposal of parts of or removed from the LuMon System
Dispose all parts of or removed from the LuMon System according to yo Follow all
local, state, or national regulations with respect to environmental protection.
WARNING
All parts of or removed from the LuMon System must be considered potentially contaminated and a source of
infection risk. Dispose of all parts removed from the device according to local regulations for medical waste.
NOTE
Dispose all parts of or removed from the LuMon System according to your institutio and/or local
regulations. Products containing electronic parts (LuMon Monitor, its power cord, belt connectors, and belts)
require special disposal. Incorrect disposal may lead to serious environmental pollution.
NOTE
Sentec is dedicated to helping protect the natural environment and to helping ensure the continued safe and
effective use of this product through proper support, maintenance, and training. Therefore, Sentec products
are designed and manufactured to comply with relevant guidelines for environmental protection. If the product
is properly operated and maintained, it presents no environmental risks. However, the product may contain
materials that could be harmful to the environment if disposed of incorrectly. Use of such materials is essential
to performing the functions of the product, and to meeting statutory and other requirements.
11.5 Instructions for repacking and shipping
Please contact your local Sentec EIT representative or Sentec AG prior to returning any material.
The following instructions apply to all shipments of the LuMon System or any part of it to your local Sentec EIT
representative or Sentec AG:
• Unless otherwise instructed please ship only the LuMon Monitor and/or belt connector, but without
any disposables such as belts or contact agent/spray.
• Devices must be properly disinfected (11.2) prior to shipment. The completed forms "Certificate of
Disinfection" and, if applicable, the "Repair Request Form" must be provided to Sentec AG per Email
and must accompany the shipment outside the package box, e.g. in a documents pouch. PDF-copies
of these forms are available from Sentec AG upon request (info-eit@Sentec.com).
• Items must be shipped in the original packaging or in other packing providing the same degree of
protection. Use sanitized packing material only.
NOTE
Any shipment to your local Sentec EIT representative or Sentec AG lacking the "Certificate of Disinfection" and,
if applicable, the "Repair Request Form" or shipments sent without written authorization by your local Sentec
EIT representative or Sentec AG may be returned unopened or disinfected at the sender nse.
NOTE
It is recommended to ship the LuMon Monitor with a fully charged battery. If possible, charge the battery
before shipping LuMon Monitor.
NOTE
Unless otherwise instructed, do NOT ship ContactAgent/LuMon Contact Spray cans.
ContactAgent/LuMon Contact Spray cans are pressurized, and therefore must be declared as "dangerous
goods", a special IATA form is required, and special packing instructions apply.