D. Connect the belt connector to the LuMon Monitor and verify that:
• the field on the right of displays a number (encodes hardware and software versions of the
connected belt connector).
• the status message connect belt is displayed on ScoutView.
If the preoperational check passed, you can initiate monitoring (8). If it failed or in case of belt connector related
error messages (6.6.2), do not use the belt connector you used during the preoperational check. In this case,
repeat the preoperational check with another belt connector.
NOTE
It is recommended to synchronize the LuMon Monitor /time (Figure 6-8, Table 6-7) with the date/time
of your institution as otherwise the date/times indicated in the trends of the LuMon Monitor or of
stored/exported data, e.g. for marked operator events (9.11), may differ from those noted in your records.
NOTE
Seasonal time changes are not automatically adjusted by the LuMon Monitor.
NOTE
EIT data
stored in the LuMon Monitor mory will be deleted when changing its date/time. No
further operator confirmation will be requested at the time of the change.
7.5 Minimum Requirements
Preoperational Minimum requirements concerning hardward, IT networks characteristics and IT security
measures:
In order to protect patient data against cyber threats, it is necessary to implement and continuously maintain
a holistic, state-of-the-art security concept. Hospitals and other heath care providers are responsible for
preventing unauthorized access to the facility systems, devices and networks when installing the LuMon
System. In case of doubt or any security issues, please consult your IT manager.