Indicates
inclination
of the patient.
-90° to -11°
Trendelen-
burg
10° to 23°
Supine,
slightly
inclined
23° to 38°
Supine,
inclined
38° to 68°
Supine,
strongly
inclined
68° to 123°
Upright
sitting or
standing
Indicates the remaining time a belt can be used.
Belt time
not elapsed, here the belt
can be used for another 71 hours
Belt time
elapsed, here since 36
hours.
Belt data not available, e.g. if no belt
is connected
Indicates the currently selected Analysis Mode (9.7).
6.6.2 Status Messages
On ScoutView (6.2), status messages are displayed in yellow between two horizontal yellow lines just
underneath the status bar.
NOTE
Status messages are only visible if ScoutView is expanded. If ScoutView is minimized only the status message
indicator line (the top of the two yellow lines) is visible (Figure 6-5).
Only one status message is displayed at a time. In case two or more conditions that require the display of a
status message occur concurrently, the LuMon Monitor displays the status message associated with the
condition that last occurred. If no status message is displayed, the field is empty, and the two yellow lines are not
displayed.
Table 6-2 lists all status messages in alphabetical order and provides for each status message a short
description.
Table 6-2: Status messages
Displays if the LuMon System has detected a belt fault.
Monitoring is not possible.
Aside from the message, the LuMon System shows the identical behavior as if a belt connector is connected
to the LuMon Monitor, but no belt is connected to the belt connector (6.6.3).
Obtain a different belt.
Displays while the LuMon System performs a system calibration (6.6.4). Usually, calibration is accomplished
within a few seconds.
Monitoring is not possible (briefly interrupted).
If calibration does not end within approximately 10 seconds, refer to 8.4 to improve the belt to skin
contact quality around the thorax.
Displays if no belt is connected to the LuMon Monitor or the connected belt is defective (6.6.3).
Monitoring is not possible.
Connect a belt or a different belt.
Is displayed if no belt connector is connected to the LuMon Monitor or the connected belt connector is
defective (6.6.3).
Monitoring is not possible.
Connect a belt connector or a different belt connector
Is displayed if the LuMon System has detected a belt connector fault.