During shipping, transport, installation, operation, if too much vibration occurs, the ferrule preload can
release and the leak appears. Other factors also affect this phenomenon like using tubing having too much
thin wall, which accelerates apparition of leaks. However, let’s not be too pessimist. Experience has shown
us that easily 95% of the fittings will work great for a very long time. We just have to be aware that
presence of leak is always possible. The only important point is to know how to check if leaks are present
and how to solve the problem.
P.S.: Servomex also manufactures a patented sampling system using a principle that eliminates almost 100
% of the leaks inside it.