Digital Weighing Indicator
SI 4100
- 45 -
6-1-7. Data Format : AD 4321 Data Transference) AD 4321 18byte Format
,
,
+,- / Including Decimal point 8byte
k
g
CR
LF
Header 1
Header 2
Data(8)
Weight
Unit
Header 1. : OL : Over Load, Under Load
ST : Display “Steady”
US : Display “Un-Steady”
Header 2. : NT : Net weight (Under Tare)
GS : Net weight (Under TARE reset)
Data Bit(Number) 2B(H) : “+” Plus
2D(H) : “-“ Minus
20(H) : “ “ Space
2E(H) : “.” Decimal Point
Unit : Kg, g, t